Biochemistry laboratory equipment: the PURELAB FLEX 3 - ULTRAPURE WATER PREPARATION SYSTEM

Purpose of equipment
A clean water system is designed to produce deionised water for further use. PURELAB flex 3 System for the production of ultrapure water from tap water. Purification principle RO, UV and cartridge (ion exchange) Type I and > III water (in tank) UV lamp 185/254nm for reduced TOC value Resistivity at 25°C 18,2MOhm-cm TOC: < 5ppb Tap water requirements Inlet water pressure 2-6 bar (recommended > 4 bar) Inlet water conductivity (recommended): < 1400uS/cm Temperature 4-40°C (recommended 10-25°C) Power supply: 100-240/56-60 V/Hz Dimensions W x D x H 236 x 470 x 900-1020 mm (with flexible arm for ultrapure water outlet) Built-in tank: 7L Tank filling rate: up to 10 l/h Ultrapure water outlet rate: up to 2 l/min with selectable water volume. Water quality and tank level display on the ultra-same water outlet arm.